ISI Web of Knowledge Take the next step  
Web of Science®
 
Previous Record (inactive) Record 1  of  1 Next Record (inactive)
Record from Web of Science®
RAPID HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ASSAY FOR TOTAL HOMOCYSTEINE LEVELS IN HUMAN SERUM
Author(s): UBBINK JB, VERMAAK WJH, BISSBORT S
Source: JOURNAL OF CHROMATOGRAPHY-BIOMEDICAL APPLICATIONS    Volume: 565    Issue: 1-2    Pages: 441-446    Published: APR 19 1991  
Times Cited: 522     References: 24     
Abstract: A rapid, isocratic high-performance liquid chromatographic (HPLC) method is described for the determination of total homocysteine levels in human serum. Prior to reversed-phase HPLC analysis, the serum thiols were derivatized with SBD-F (ammonium 7-fluorobenzo-2-oxa-1,3-diazole-4-sulphonate), a thiol-specific fluorogenic probe which is commercially available. Retention of SBD-homocysteine was sensitive to pH, and a mobile phase pH of 2.1 ensured baseline separation of serum thiols within 6 min. The method is simple, sensitive, reproducible (between-run coefficient of variation of 6.6%) and very suitable for routine determination of serum homocysteine levels in a clinical pathology laboratory.
Document Type: Note
Language: English
Reprint Address: UBBINK, JB (reprint author), UNIV PRETORIA, FAC MED, DEPT CHEM PATHOL, POB 2034, PRETORIA 0001, SOUTH AFRICA
Publisher: ELSEVIER SCIENCE BV, PO BOX 211, 1000 AE AMSTERDAM, NETHERLANDS
Subject Category: Chemistry, Analytical
IDS Number: FK218
ISSN: 0378-4347